Table 1. Parameters and Error Indication
Parameter Min Max Error indication for illegal
values.
(x is the channel number)
M 1 59,900 Mx Err
DURATION 40 msec 3999 sec Dx Err
DELAY 100 msec
Delay> Duration/10,000
3999 sec Lx
Lx
Err
Err
INTERVAL
a)Normal usage:
b)Train Mode:
c)If the channel
internally activates
other channels:
60 msec
Interval> Duration + 9 msec
Interval> Duration + 59 msec
Interval >500 msec
3999 sec Ix
Rx
Tx
Cx
Err
Err (Rate err)
Err (Train err)
Err (Connect err)
Example: The message ‘R7 Err’ indicates ‘Rate error’ on channel #7. The reason is that the channel is
now running in FREE-RUN, GATE or TRAIN mode and the interval is shorter than the duration or at
least
INTERVAL < DURATION + 9 msec.
In order to remove the R7 Err message you should do one of the following:
One. increase the interval time of channel #7 or
Two. decrease the duration time of channel #7 or
Three. change the mode of channel #7 to TRIG, DC or OFF.
Please note that there is no delay in the TRAIN mode. The channel delivers the train of pulses
immediately after receiving the trigger input. In order to obtain a delay time, you should use an
intermediate channel in the TRIG mode.